1 fluorescein
1 fluorescein kdorfman Mon, 08/19/2013 - 17:091.1
1.1 kdorfman Mon, 08/19/2013 - 17:14Materials
- one balance per table (borrow from Bill Schmitt)
- weighboats
- non-sterile tips
- Mats
- Gloves
- small glass culture tubes (~15/pair)
- rack for tubes
beaker for water
blue flashlights & yellow filters
Computers and Plate Readers
- Check program Repeatability (Abs 485 nm; layout A1-B12)
- Wahoo folders must be named Tues01, Tues02, ...Tues12 and Thur01, Thur02 ... Thur12, even though the user names in the Optima program are Tue01 and Thu01, etc.
- Import program to groups
- Check network connections (in “my computer” connect the Z drive to Wahoo as kdorfman, kd@ISB)
10mM NaOH
MW = ~40
10mM = 0.01M = 0.4g/L
1 mM Fluorescein
in 10 mM NaOH
It fades with time, so check the concentration of the old stocks.
MW = 332
1mM = 0.332g/L = 0.033g/100 mL
A = Ecl
Molar extinction coefficient of fluorescein at 485nm = 50358/mol.cm
so Absorbance 10^-5M = 0.50358
(1/100 of a 1mM solution =10^-5M)
1.2
1.2 kdorfman Mon, 08/19/2013 - 17:18Absorbance of Fluorescein
Per pair:
- Black 96 well plate (not necessarily sterile, but clean)
- 1 mM fluorescein (~ 1 mL)
- 10 mM NaOH (~ 4.5 mL)
- 2 2-mL tubes to make 2 mL of 10µM fluorescein (each partner does one)
Computers and Plate Readers
- Check program Repeatability (Abs 485 nm; layout A1-B12, volume 150 µL)
- Check program FL-Abs-by-2 (Abs 485 nm; layout C1-2, D1-12, volume 150 µL)
- Wahoo folders must be named Tues01, Tues02, ...Tues12 and Thur01, Thur02 ... Thur12, even though the user names in the Optima program are Tue01 and Thu01, etc.
- Import program to groups
- Check network connections (in “my computer” connect the Z drive to Wahoo as kdorfman, kd@hh3)
solutions
solutions kdorfman Fri, 09/06/2013 - 18:471 mM Fluorescein
in 10 mM NaOH
It fades with time, so check the concentration of the old stocks.
MW = 332
1mM = 0.332g/L = 0.033g/100 mL
A = Ecl
Molar extinction coefficient of fluorescein at 485nm = 50358/mol.cm
so Absorbance 10^-5M = 0.50358
(1/100 of a 1mM solution =10^-5M)
10mM NaOH
MW = ~40
10mM = 0.01M = 0.4g/L
1.3
1.3 kdorfman Mon, 08/19/2013 - 17:18Corrections
Read the plate only once
Pipet 150 µL of each unknown in E3 - 12 as needed
Programs
- Serial2&5_Abs
- Serial2&5_FL
- E1 &E2: blank E3-E12 for unknowns
- F1 - F12: dilute by 2
- G1 - G12: dilute by 5
Materials:
- ~1 mL 1mM fluorescein (try to make it right via Abs)
- E = 50358/M
- A = 50358/M * C
- at 10 µM, A = 0.5038
- dilute 1 mM 1/100 to make 10 µM, take Abs
- 5 mL 10 mM NaOH
- black plate from last time (or new one if old one has no clean wells)
Conversion factors
Abs = 59.7*(mM)-0.341 up to mM=0.03125
FL = 10^6 * mM +173.06 up to mM = 0.008
FL = 2 X 10^6 * mM + 19.624 up to mM = 0.00195
Unknowns
- Make dilution series from the "1mM" stock.
- find lowest useful concentration for Abs. Dilute slightly for unknown 1 - must be done with FL. Abs is bad below 0.0005 and meaningless below 5 x 10^-7mM
- find a concentration too high for fluorescence, but OK for abs, e.g., 0.05.
- find a concentration too high for either. must be >0.125
2013:
label | ~mM | Abs | FL |
---|---|---|---|
P | 0.0001 | ~0 | 210 |
Q | 0.015 | 0.8699 | 16496 |
R | 0.1 | 4.955 | 26780 |
2012
2012 kdorfman Mon, 08/19/2013 - 17:121.1: Fluorescein
1.1: Fluorescein kdorfman Mon, 11/14/2011 - 21:27Fluorescein Absorbance and Fluorescence
Materials for 1.1
Materials for 1.1 kdorfman Mon, 11/14/2011 - 21:32Materials
- 1 black-walled plate per pair
- non-sterile tips
- Mats
- Gloves
- Serial dilution by 5 for demonstration purposes.
- First tube saturated
- Last tube blank.
- blue flashlights & yellow filters
Computers and Plate Readers
- Check program Repeatability (Abs 485 nm; layout A1-B12)
- Wahoo folders must be named Tues01, Tues02, ...Tues12 and Thur01, Thur02 ... Thur12, even though the user names in the Optima program are Tue01 and Thu01, etc.
- Import program to groups
- Check network connections (in “my computer” connect the Z drive to Wahoo as kdorfman, kd@hh3)
Reagents for 1.1
Reagents for 1.1 kdorfman Mon, 11/14/2011 - 21:2810mM NaOH
MW = ~40
10mM = 0.01M = 0.4g/L
Need:
- 25 tubes of 10 mL each = 250 mL NaOH
13 tubes of 2 mL each = 26 mL fluorescein A
13 tubes of 2 mL each = 26 mL fluorescein B
= total of 300 mL NaOH needed
1 mM Fluorescein
in 10 mM NaOH
It fades with time, so check the concentration of the old stocks.
MW = 332
1mM = 0.332g/L = 0.033g/100 mL
A = Ecl
Molar extinction coefficient of fluorescein at 485nm = 50358/mol.cm
so Absorbance 10^-5M = 0.50358
(1/100 of a 1mM solution =10^-5M)
Make A & B slightly different.
(Different enough so that they are different by the t-test. Make one above and one below the “actual” value.)
Be sure to measure a blank. In 2012, the blank was ~0.25, so Abs readings should be ~0.75
1.2: Fluorescein II
1.2: Fluorescein II kdorfman Mon, 11/14/2011 - 21:40Absorbance & Fluorescence of Fluorescein, part 2
Unknowns
One too dilute to measure with abs (e.g., 0.0003mM; A3x10^-7M = 0.015)
One too concentrated to measure with Fl (e.g., above 0.01mM)
One too concentrated for either, e.g., 0.12 mM
Tell instructors the Abs and Fl of each, w/gain adjusted for the 1 mM solution
Give students:
~6 mL 10mM NaOH
~1 mL 1mM fluorescein. (Make sure Abs is correct)
Make dilution series
Check programs for correct wells.
Get a laptop for the TA to demonstrate stupid Excel tricks.
2012 unknowns
2012 unknowns kdorfman Tue, 09/18/2012 - 17:51Unknown | ~ mM | ~Abs | Abs blank | ~ Fl | Fl blank |
---|---|---|---|---|---|
P | 0.25 | 5.745 | 0.023 | 38000 | 19 |
Q | 0.00024 | 0.039 | 0.023 | 679 | 19 |
R | 0.0625 | 4.171 | 0.023 | 32000 | 19 |