Materials
- one balance per table (borrow from Bill Schmitt)
- weighboats
- non-sterile tips
- Mats
- Gloves
- small glass culture tubes (~15/pair)
- rack for tubes
beaker for water
blue flashlights & yellow filters
Computers and Plate Readers
- Check program Repeatability (Abs 485 nm; layout A1-B12)
- Wahoo folders must be named Tues01, Tues02, ...Tues12 and Thur01, Thur02 ... Thur12, even though the user names in the Optima program are Tue01 and Thu01, etc.
- Import program to groups
- Check network connections (in “my computer” connect the Z drive to Wahoo as kdorfman, kd@ISB)
10mM NaOH
MW = ~40
10mM = 0.01M = 0.4g/L
1 mM Fluorescein
in 10 mM NaOH
It fades with time, so check the concentration of the old stocks.
MW = 332
1mM = 0.332g/L = 0.033g/100 mL
A = Ecl
Molar extinction coefficient of fluorescein at 485nm = 50358/mol.cm
so Absorbance 10^-5M = 0.50358
(1/100 of a 1mM solution =10^-5M)